|
Volume 1, Article 34
A. P. Kovarsky, V. S. Strykanov
Surface Diagnostics Lab., Mekhanobr-Analyt Co.
In this study we optimized the analysis conditions to determine detection limits of impurities of interest for controlling the conductivity of GaN films. In-depth analyses of contaminants such as H, C, O is also shown.
A Cameca IMS 4f instrument was used. The analysis conditions were defined from empirically established principles for AIIIBV semiconductors [4]. We selected 5.5 keV O2 + primary ions to analyze Mg, Al and In; positive secondary ions were recorded. Cs+ primary ions (E = 5.5 keV) were used for Zn and Cd and CsX+ ions (X= Zn, Cd) were detected. The Si, H, C, O signals were obtained using negative secondary ion registration. Primary current densities of 100 - 1000 μA/cm2 were employed. The sputter rate for GaN was 2 - 10 Å/sec for the various of primary ions.
To eliminate a crater edge effects we chose the field and contrast diaphragms so that only a 60 μm diameter area centered in the 250 μm × 250 μm crater was analyzed.
Some samples showed initial Mg matrix contamination which could be in the form of local inclusions revealed in our ion images of analyzed area. The detection limit for Mg presented in figure 1 corresponds to the "pure" matrix background level.
Local inclusions were more pronounced when we analyzed for Al (figure 2). We do not discuss the nature of the inclusions in this paper, but we believe their presence explains the high background signal level. Therefore, we believe that it is reasonable to expect that a 1015 atoms/cm3 detection limit for Al (dotted line, figure 1) would be possible in the "pure" matrix case.
The detection limit of In is defined by the superposition of the matrix complex ions, such as (71Ga14N216O)+, (71Ga28Si16O)+ or (71Ga27Al16O1H)+, and the 115In+ signal. To minimize the influence of the complex ions it is possible to change the secondary accelerating voltage (offset ~ 90 V). This method allows one to discriminate complex ions by means of an energetic slit because they have a more narrow energetic distribution in comparison to monatomic ions. When complex ions are minimized, the In detection limit decreases by one order or magnitude (solid squares, figure 1).
The determination for Si (figure 3) was realized by analyzing negative secondary ions (28Si-) with the aid of the high mass resolution mode (ΔM/M = 3000). The N2- matrix signal was taken as the reference. This is a traditional method for optimal Si determination in AIIIBV semiconductors [6] [7]. The background signal appeared to be one order of magnitude above the detector background due to volume contamination of Si in the GaN matrix. The detection limit is actually much lower than the 10.5 x 1016 atoms/cm3 represented by the dotted line in figure 3.
It must be mentioned that the high resolution mode is difficult to realize when one must record concentration profiles of several elements simultaneously. In-depth analysis of Si and Mg in GaN can serve as an example. We estimated the Si detection limit at optimal Mg analyzing conditions (O2+ primary beam and 44SiO+ registration) to be only 1018 atoms/cm3. Hence, if it is not necessary to detect very low Si concentrations, it is possible in normal mode operation (ΔM/M ~ 300) to determine the position of a GaN pn junction directly.
The reason for the high background signal levels for the gaseous impurities is clearly illustrated in figure 5 which shows ion images of the analyzed areas. The images were obtained under conditions identical to those of the analyses over a 150 μm diamter. Segregation of these impurities to the crystallite boundaries is clearly observed. Carbon forms also incorporations with lack of H and O. We conclude that these impurities are incorporated during the crystal growth. We do not believe that it would be a problem to achieve detection limits for these elements of ~1016 atoms/cm3 because the background level corresponding to the impurity concentration in the matrix is three orders of magnitude higher than the ion detector noise level.
Unfortunately, attempts to achieve low detection limits for N, C and O have failed. We postulate that the measured minimum sensitivities of Cmin ≈ 1018 - 1019 atoms/cm3 are caused by the segregation of these impurities on crystallite boundaries.
Finally, we mentioned that our data collection was complicated by sample charging during the ion bombardment and methods must be employed to minimize this effect.
[1] H. Morkoc, S. Strite, G. B. Gao, M. E. Lin, B. Sverdlov, M. Burns , J. Appl. Phys. 76, 1363-1398 (1994).
[2] M. Gauneau, R. Chapling, A. Rupert, M. Salvi, B. Desconts, J. Vac. Sci. Technol. A 8, 4039 (1990).
[3]D. P. Leta, G. H. Morrison, “Ion Implanted Standards for Secondary Ion Mass Spectrometry Determination of the 1a-7a Elements in Semiconducting Matrix”, Analytical Chemistry 52 (3), 514 (1980)
[4]R. G. Wilson, F. A. Stevie, Ch. W. Magee, “Secondary Ion Mass Spectrometry. A Practical Handbook For Depth Profiling And Bulk Impurity Analysis” John Wiley and Sons, New York, (1989)
[5] M. Gaunean, R. Chaplain, A. Rupert, J. Phys. (Paris) 45, 119 (1984).
[6] Y. Homma, Y. Ishii, J. Appl. Phys. 56, 1892 (1984).
[7] K. Inoue, H. Sakaki, J. Yoshina, Y. Yoshioka, Appl. Phys. Lett. 46, 973 (1985).
Table I
Ion implantation data for the GaN samples
| Sample# | Isotope | Dose [cm-2] | Energy [keV] |
| 43 | 24Mg | 3e15 | 200 |
| 63 | 24Mg | 1e15 | 200 |
| 65 | 24Mg | 1e15 | 200 |
| 62 | 27Al | 1e15 | 200 |
| 64 | 27Al | 1e15 | 200 |
| 15 | 115In | 5e14 | 100 |
| 24 | 28Si | 1e15 | 100 |
| 41 | 28Si | 5e15 | 200 |
| 61 | 28Si | 1e15 | 200 |
| 50 | 64Zn | 6e14 | 200 |
| 53 | 114Cd | 3e15 | 200 |
| 60 | 114Cd | 3e14 | 200 |
| 31 | 16O | 1e15 | 100 |
| 23 | 12C | 1e15 | 100 |
| 32 | 12C | 1e15 | 100 |
| 33 | 1H | 1e15 | 50 |
| Mg | Al | Si | Zn | Cd | In | H | C | O | |
| Imax /Imin | 5e4 | 1e4 | 1e3 | 1e3 | 1e3 | 1e4 | 5-7 | 10 | 10 |
| Cmin, at./cm3 | 2e15 | 2e16 | 1e17 | 3e16 | 5e16 | 1e16 | 1e19 | 3e18 | 5e18 |
| Figure 1. Calibration characteristics of Mg, Al and In impurities in GaN. |
Figure 2. Ion image (27Al+) of Al contamination in the GaN-matrix (image field is 150μm). |
| Figure 3. Calibration characteristics of Cd,Zn and Si impurities in GaN. |
| Figure 4. Depth profiles of H, C and O in implanted samples. |
Figure 5. Ion images of H, C and O contaminations in the GaN matrix (a -GaN-, b - H-, c - CN-, d - O-). |
© 1996-1997 The Materials Research Society
|